How to Start Your Clean-Trace ATP-meter

Have you just bought a new Clean-Trace but find it difficult to turn on or off? Follow our tips below to avoid unnecessary problems.

Tips for starting and shutting down:

  • Charge the meter fully: Make sure the meter is fully charged (at least 2 hours) before use. You can charge it via a USB port on a computer, but the fastest way is to use ATP own charger.
  • Start the timer: Press and hold the start button (A) for at least 3 seconds. Startup includes an automatic function check that takes about 70 seconds.
  • Sleep mode:
    1. If the meter is not used for 5 minutes, the screen turns off and the meter enters sleep mode. There is no indicator light to show this, which may give the impression that it is turned off. After 2 hours of inactivity, the meter turns off completely.
    2. Press the mouse button (B) to wake the instrument from sleep mode.
    3. Not sure if the meter is in sleep mode? Get into the habit of always starting by pressing the mouse button (B) to avoid unnecessary restarts.
  • Adjust the sleep timer: Do you think 5 minutes is too short? Log in to the meter and change the sleep timer to 10 or 30 minutes in the settings.
  • Turn off the meter:
    1. Press and hold the start button (A) for at least 3 seconds until a pop-up window appears.
    2. Tap “Shut Down” on the screen .
      NOTE: If you press and hold the button for a shorter time, only the screen saver (black screen) will be activated, not the shutdown pop-up window. Click the mouse button (B) to wake up the screen! 
Listing
If you press on/off/on/off, etc., repeatedly while the screen is off, there is a risk of crashing the system.Werecommend that you practice the tips above to avoid this problem. If you have any further questions about your ATP-meter free to call or email our product manager, Frank Axelsson, at 0707 912103.