Have you just bought a new Clean-Trace but find it difficult to turn on or off? Follow our tips below to avoid unnecessary problems.

Tips for starting and shutting down:
- Charge the meter fully: Make sure the meter is fully charged (at least 2 hours) before use. You can charge it via a USB port on a computer, but the fastest way is to use ATP own charger.
- Start the timer: Press and hold the start button (A) for at least 3 seconds. Startup includes an automatic function check that takes about 70 seconds.
- Sleep mode:
- If the meter is not used for 5 minutes, the screen turns off and the meter enters sleep mode. There is no indicator light to show this, which may give the impression that it is turned off. After 2 hours of inactivity, the meter turns off completely.
- Press the mouse button (B) to wake the instrument from sleep mode.
- Not sure if the meter is in sleep mode? Get into the habit of always starting by pressing the mouse button (B) to avoid unnecessary restarts.
- Adjust the sleep timer: Do you think 5 minutes is too short? Log in to the meter and change the sleep timer to 10 or 30 minutes in the settings.
- Turn off the meter:
- Press and hold the start button (A) for at least 3 seconds until a pop-up window appears.
- Tap “Shut Down” on the screen .
NOTE: If you press and hold the button for a shorter time, only the screen saver (black screen) will be activated, not the shutdown pop-up window. Click the mouse button (B) to wake up the screen!
Listing
If you press on/off/on/off, etc., repeatedly while the screen is off, there is a risk of crashing the system.Werecommend that you practice the tips above to avoid this problem. If you have any further questions about your ATP-meter free to call or email our product manager, Frank Axelsson, at 0707 912103.