LuciPac® A3 Surface – dry
LuciPac® A3 Surface is an all-in-one test for quickly verifying the cleanliness of surfaces. The test uses dry swabs to measure ATP, as well as the breakdown products ADP and AMP, using Lumitester, and provides results within 10 seconds.
Description
LuciPac A3 Surface is used to check whether surfaces have been cleaned sufficiently. The swabs are dry and must be moistened before use on dry surfaces.
Each sampling unit contains a sampling swab, extraction solution, and measurement reagent. After sampling, the test is activated and placed in a compatible Lumitester Smart or PD-30 meter. The result is displayed in RLU within 10 seconds.
This product is intended for routine hygiene inspections in, for example, the food industry, commercial kitchens, healthcare facilities, laboratories, and professional cleaning services.
Features
- Dry swab – NOTE: Must be moistened before use on dry surfaces
- Simple all-in-one device with sample collection and reagents
- Measures ATP, ADP, and AMP
- Fast measurement results within 10 seconds
- Compatible with Lumitester Smart and Lumitester PD-30
- AOAC Performance Tested Methods℠ PTM 112201
LuciPac® A3 Surface – Pre-Moistened features a swab that is pre-moistened and ready for immediate use on both dry and damp surfaces. Pre-moistened swabs provide more standardized (reproducible) sampling because the moisture level is more consistent, and there is no need to re-moisten the swab with clean water before sampling dry surfaces.
LuciPac® A3 Surface – Dry features a swab that can be used directly on damp surfaces or moistened separately before sampling on dry surfaces. The main advantage of dry swabs is their longer shelf life: approximately 15 months from the date of manufacture, compared to 12 months for the pre-moistened version.
We recommend choosingLuciPac® A3 Surface – Pre-moistened first and foremost,as it is easier to use and more standardized. However, the dry version is best if a longer shelf life is the top priority.
Lightly moisten the swab tip with clean tap water before sampling dry surfaces. The swab should be evenly moistened but not soaked or dripping, as too much water can impair sample collection and make the results less reproducible.
Our recommendation:
- Let the cold water run for a few seconds.
- Moisten the swab immediately—do not use a reused cup.
- Regularly measure a " water blank," that is, a moistened swab without a surface sample.
- If the blank value is more than approximately 10–20 RLU, check the water and your procedure.
- For maximum reproducibility: choose the pre-moistened A3 Surface fromLuciPac , where the moistening is standardized.
Instructional video:
More information
| Parameter | ATP |
|---|---|
| Result type | Quantitative |
| Test type | Surface swab – round |
| Number | |
| Storage temperature | 2–8°C |
| Sustainability | Batch-dependent, as per packaging |
| Other specifications | Test type: dry swab for surface testing in hygiene inspections |
This product is usually in stock; delivery takes approximately 1–3 days.
Please contact us for stock availability, delivery times, a quote, or assistance in choosing the right product.
Call or email our product experts—or use our contact form:
Tel: +46 18-470 44 90
Email: [email protected]
Contact us
Related products
You might also like …
Item No.: 60365
LuciPac® A3 Surface – Pre-moistened
Quantity: 100 pcs. per packSign in / register for pricing Read moreItem No.: 60367
Item No.: 61234